***************************************************************** Subject: G130x-004 Supersedes: None. Product Number G1304A, G1305A, G1307A, G1309A (revision A.02.00) Serial Number: n.a. Headline Software patch available for aborted sequences due to a VWD signal error. Parts required: see below. Situation: Due to an inconsistency in the way the HPLC ChemStations (DOS Series), revision A.02.00 decode the error bits generated by the HP 79855 C (VWD), sequences may be aborted after a signal overflow in the VWD. Solution/Action: A software patch (revision number A.02.01) is available from the analytical response centers that solves this problem. The patch consists of a single executable file that has to be copied to the system directory of the HPLC ChemStation e.g c:\hpchem\lc3dexe This defect will be fixed in the next release (A.02.02), scheduled for september, 1993. Date: August 13, 1993 ----------------------------------------------------------------- Modification Available ------------------------ Reason: Performance Enhancement __ -or- Service/Reliability Enhancement XX Location Category: Customer Installable XX On-Site __ HP location __ Available until: release A.02.02 Author/Entity: Wolfgang Winter, B400 ----------------------------------------------------------------- *****************************************************************